logo

Determination of Orthophosphate in Ammonium Polyphosphate by Ion Chromatography for Fertilizer Quality Control

2026-11-11

최신 회사 사례 Determination of Orthophosphate in Ammonium Polyphosphate by Ion Chromatography for Fertilizer Quality Control
사례 세부정보

Background

Ammonium polyphosphate (APP), also called polymerized or condensed ammonium phosphate, is widely used as a compound fertilizer. Low-polymerization grades are highly water-soluble, carry high nitrogen and phosphorus content, and release their phosphorus slowly. Compared with conventional phosphorus fertilizers, polymerized phosphate offers better solubility, a longer lasting effect, higher nutrient use efficiency, greater safety for soil and crops, and support for balanced fertilization.

Because the slow-release behaviour of APP depends on how much of its phosphorus is still in polymerized form, the polymerization rate — the mass fraction of polymerized phosphate in total phosphorus — is a key quality indicator. The Chinese chemical industry standard GB/T 5939-2021 (Fertilizer grade ammonium polyphosphate) describes two determination routes: an ion exchange resin method and an ion chromatography method. Both separate orthophosphate from polymerized phosphate on a resin or IC column before measuring the orthophosphate content. This case reports the results obtained with a Wayeal ion chromatograph applying the IC approach.

Instrument Configuration

ItemDescriptionQty
Ion chromatographIC6600 series with conductivity detector1
AutosamplerAS28001
SoftwareSmartLab CDS 2.0 chromatography workstation1
Analytical columnHS-5A-P3, 4.0 × 250 mm1
Consumables0.22 µm disposable syringe filters; orthophosphate standard 1000 mg/L—

Sample Preparation and Chromatographic Conditions

Approximately 0.02 g of ammonium polyphosphate sample is weighed, dissolved in ultrapure water and made up to 100 mL. The solution is passed through a 0.22 µm disposable syringe filter and injected directly — no derivatisation or complex clean-up is required.

ParameterSetting
ColumnHS-5A-P3, 4.0 × 250 mm
EluentIsocratic, 40 mmol/L KOH
Flow rate1.0 mL/min
Run time45 min
Injection volume25 µL
Column temperature30 °C
Cell temperature30 °C
Suppressor current120 mA

Results

The determination of orthophosphate in the ammonium polyphosphate sample was completed within 45 minutes, with well-shaped peaks and good response. A blank injection showed no orthophosphate (0 mg/L), confirming a clean baseline.

Linearity: a series of diluted orthophosphate standards was prepared. The linear range was 10–50 mg/L with a correlation coefficient R² of 0.99972, and the deviation from the known concentrations stayed within the maximum permitted tolerance.

Sample results: three parallel determinations were run on a sample with 60.48% available phosphorus.

SampleWeight (g)Volume (mL)Measured (mg/L)Orthophosphate (%)
Parallel 10.021510020.8799.71
Parallel 20.022710020.9229.22
Parallel 30.023910021.9269.17
Average———9.367

Taking the average of the three parallels, the orthophosphate content was 9.367%. The polymerization rate — the difference between available phosphorus and orthophosphate relative to available phosphorus — was 84.5% by parallel average and 84.8% for the reported sample result, showing good repeatability across injections.

Conclusion

The Wayeal ion chromatograph, configured with a conductivity detector, AS2800 autosampler and HS-5A-P3 column, provides a precise and efficient route for monitoring orthophosphate in ammonium polyphosphate. The method supports fertilizer-grade product quality control, and the polymerization rate it delivers helps manufacturers assess hydrolysis behaviour in soil, evaluate phosphorus release capability, optimise fertilization programmes and improve overall fertilizer performance.